P95 Manual: Ivoclar Programat
Check the firing temperature (B), heat-up rate (t), and holding time (H).
is a microprocessor-controlled, fully automatic vacuum ceramic furnace designed for high-accuracy dental ceramic firing. Ivoclar Programat P95 Manual
is a microprocessor-controlled, fully automatic vacuum ceramic furnace designed for high-precision dental applications. It features a numeric keypad for data entry and supports up to , making it a reliable tool for consistent porcelain results. Key Features and Specifications Capacity: 99 programmable firing slots. Check the firing temperature (B), heat-up rate (t),
| Parameter | Range | Default | Clinical Impact | | :--- | :--- | :--- | :--- | | Heating rate | 1 – 100°C/min | 10°C/min | Too fast cracks green-state ZrO₂. | | Final temp | 800 – 1600°C | 1530°C | ±10°C changes translucency. | | Hold time | 0 – 999 min | 90 min | Insufficient hold = low density. | | Cooling rate (R1/R2) | 1 – 40°C/min | 12°C/min | Too fast = crown cracks. | | Vacuum on/off | On/Off | Off (Sinter) | On = oxygen vacancies (gray). | It features a numeric keypad for data entry
Before operating the Programat P95, the manual emphasizes strict adherence to installation protocols to ensure safety and longevity of the unit.
Here is what you will find in the official documentation and why each section matters.
Avoid placing the furnace near air conditioning vents or open windows to prevent temperature fluctuations. Connecting the Vacuum Pump